CAPEL-115 Poptable Capillary Electrophoresis System from Lumex Instruments
  • CAPEL-115 Poptable Capillary Electrophoresis System from Lumex Instruments

CAPEL-115 Portable Capillary Electrophoresis System



Logo Lumex Instruments

Reference : CAPEL-115



Capel-115 is a portable capillary electrophoresis (CE) system intended for determination of a wide range of analytes in various types of samples, including water, soil, food, and fodders using main modes of separations available in capillary electrophoresis.



Features & Advantages 

  • Portable, light, and field deployable
  • Small footprint
  • High separation efficiency
  • Multiple components determination in one run
  • Suitable for various types of analytes from small ions to macromolecules
  • Liquid cooling of capillary ensures high reproducibility
  • Fast analysis (most analyses take 5–15 minutes)
  • Simple sample pre-treatment (just dilute and filter)
  • Low analysis cost (nanoliters of reagents, only low-cost consumables are needed)
  • Power supply unit equipped with a built-in rechargeable battery and AC power cord
  • IQ/OQ protocols





Principle of Operation

High performance capillary electrophoresis is based on the differential migration of components of aqueous samples within a narrow fused silica capillary driven by an electric field. Separated solutes are quantitatively detected at the capillary outlet by a high sensitive optical system based on direct or indirect UV absorbance.



Applications

  • Petrochemical industry
  • Environmental analysis
  • Food and beverages production
  • Animal feed industry



Equipment and Options

  • Capillary electrophoresis system Capel-115
  • Spare capillary cassette
  • Elforun software package
  • Kits & sets for analysis (by request)
  • Service agreement (by request)



Technical Specifications

Detection wavelength :

190–380 nm

Analysis :

Constant voltage, from –30 kV to +30 kV in 1 kV steps

Automatic (electronic) polarity switch

Current 0–300 μА

Pressure, up to 100 mbar

Programmable changing of wavelength, pressure and voltage during analysis

Injection :

By voltage, from –30 kV to +30 kV in 1 kV steps;

by pressure, from –100 to 100 mbar in 1 mbar steps

Rinsing :

By pressure, 500–2000 mbar in 1 mbar steps

Temperature control of capillary :

Liquid thermostating, from -10 up to+ 30 °C from ambient temperature, ±0.1°C

Sampler :

Autosampler for 10 vials (standard 1.5 mL)

Power requirements :

100–240 VAC, 50/60 Hz or built-in battery 15 VDC

Environmental conditions :

ambient air temperature, 10 – 35 °C

relative air humidity at 30 °C, not more 80 %

atmospheric pressure, 84 – 106.7 kPa

Time of continuous operation with built-in battery :

5 h

Dimensions / Weight :

290х385х460 mm, 15 kg



CAPEL-115

Specific References

Download

Lumex_CAPEL 115

Datasheet Capel-115 Portable capillary electrophoresis system from Lumex Instruments

Download (507.11KB)